Here, we will show some usual questions our customers reflected for your reference. For this questions, it is very easy to operate even though you are not a very professional personal.
Following questions are of Step Setting Problem, Data problem, Channel Problem, Software Problem, Connection Problem,etc. More kinds of questions are waiting for your reflect. If you can not find the answer which can solve your problem, please try (618)233-8866 or contact us at jessie@newarebattery.comFirstly, check the connection of your tester and computer is right. Then check whether you log in the system or not. Only logging in the system, can we start the test. User: admin Password: neware
If you never changed the login password, and it always show password incorrect when you login in. Please unload the software and install again.
When you open your NDA data of one channel, the data is incomplete because of some reason.
What you can do to view the complete data. Click here to see the complete solution
Firstly, check the unit ID of your old tester and new one to see whether the ID is same or not. If the same, you need to change the new tester unit ID. Please check the information label pasted on your tester. From the equipment label, you can judge your tester belongs to 4000 series or 3000 series.
For 4000 series, click here to get the method of changing unit ID of tester. And 3000 series, click here.
When you export your data to Excel, you find that the current is quite different from real current in constant charge period, and some time it is even 0Ah. But in BTSDA, the currents shows normally, why will it happen and how to solve it? Click here
Generally, when you are asking for the after-sale service, you need to provide the picture of information label, then we can know your tester information. Click here to know the meaning of the information label.
For our 4000 series tester, only the tester whose voltage is lower than 20V and current upon 50A can support channel parallel.
For the first time to use our tester, we have a guidance file about the connection between computer and devices. Click here to get it.
Generally, the cut-off current is 1% of the battery whole capacity
For 4000 series, only the tester whose voltage is lower than 20V and current is more than 50A can support channel parallel. Two adjacent channels on the same tester can be connected in parallel. The positive current cables and positive voltage cables of two channels need to be connected with battery anode , and negative cables connected battery cathode. Software operation please click here.
If you did not change the password and show password incorrecy, please try to unload the client software and reinstall it again.
In this case, it is possible that the physical and logical order of the control unit and channels did not match, you can restart the control unit, search for tester channels again, and then remap.
Perhaps deleted the configuration file in the client software or server by mistake, it is recommended to update the latest client software.
The default cycle quantity displaying on BTSDA is 20. If you require to show more. Please double click the yellow graduated scale located in the bottom of software, and enter the cycles No. you want to display. Click here to get more details.
It is recommend to close the antivirus software.
The sampling time can be set according to your real needs. If it is just grading test, 10s is Okay.
To set different sampling in a test, you are supposed to choose “ Professional Settings” not “ Standard Settings” before editing the step. Then for each step you can enter the sampling time.
If you need to upgrade the software, first make sure previous test data and steps settings have been saved. The previous data and test cannot be continued after the upgrade program.
The large amount of data is because the channel sampling time is too small. You can set a sampling time of 10s or 30s to reduce the number of data records.